Table 7- 5 Input/Output Wiring (Tb2) - 7Me39400Al03 Expanded I/O Module - Siemens SITRANS FUH1010 Product Instruction Manual

Liquid ultrasonic flowmeter / interface detector (ifd)
Hide thumbs Also See for SITRANS FUH1010:
Table of Contents

Advertisement

Functions
7.3 Span Data
Table 7- 5
Input/Output Wiring (TB2) - 7ME39400AL03 Expanded I/O Module
Pin#
Signal
12
PG4
POS [+] Total TTL
11
PG3
POS [+] Total OC
10
PG2
NEG [-] Total TTL)
9
PG1
NEG [-] Total OC
Figure 7-1
Note
TB2-9 and TB2-11 are Open Collector Outputs that require external pull-up resistors for
operation. See table for External Supply Voltage and suggested resistor value and ratings.
Maximum current into the transistor is 100mA. Maximum Voltage is +36 VDC.
NOTICE
Transistor Damage
Negative voltages with respect to ground will permanently damage transistors.
Use caution when applying power to circuit boards.
96
Definition
0-5000 Hz frequency output ,
assignable
Use reference ground for returns (TB2-2 and TB2-4)
+5V
PG4
DPGEN_2-Ø2
2
4
TTL/CMOS
PG3
DPGEN_1-Ø2
1
+5V
PG2
DPGEN_2-Ø1
2
4
TTL/CMOS
PG1
DPGEN_1-Ø1
1
PGEN Wiring Diagram
Description
POS [+] Total TTL
POS [+] Total OC
NEG [-] Total TTL
NEG [-] Total OC
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA
Function
Dual/Quad Path Only

Hide quick links:

Advertisement