Operation
Manual Operation
2
Select the desired rotor, and press the
• The Main screen returns to view, and the selected rotor name and serial number appear on
the
a. If you were about to start a run and were required to select a rotor, the run starts
automatically when the Main screen returns to view.
3
To convert RPM to RCF for the selected rotor, press the
• The conversion is calculated automatically, and the set RCF value appears above the
Speed
— "

Entering Run Speed

Run speed ranges from 5000 to 150,000 RPM. If no new speed is entered, the ultracentrifuge
automatically selects the last entered speed. The last three digits of the run speed are fixed as zeros
and cannot be changed. For example, to enter a run speed of 100,000 RPM, type 100 on the keypad
display.
NOTE
To enter or view an RCF value, you must select the rotor first. For information on selecting a rotor, see
Selecting a
3-8
button.
ROTOR
display.
" displays in green above the horizontal line on the
RCF
Rotor.
button to accept.
OK
RPM/RCF
button.
button.
RPM/RCF