Test Connections; Testing Power Input Cables And Circuit Breakers - Emerson Rosemount CT5800 Quick Start Manual

Continuous gas analyzer
Hide thumbs Also See for Rosemount CT5800:
Table of Contents

Advertisement

Install
2.4

Test connections

Before sealing the cable glands, it is vital to test the new connections in order that any
faults can be remedied before the cables are permanently sealed.
WARNING!
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
The cable glands used in this analyzer are only suitable for use in areas with a low risk of
mechanical damage and must be suitably protected.
Failure to observe this warning could cause an explosion or potentially hazardous situation
which, if not avoided, may cause death or personal injury.
The cable glands can be temporarily sealed with a suitable sealant to allow testing.
The tests described below are a basic check to prove function, rather than a test of sensor
capability.
The analyzer should be started according to the startup procedure in
WARNING!
INSTALLER RESPONSIBILITY
In accordance with IEC 60079-0: 2011 Clause 30 + CSA report.
It is your responsibility to connect the analyzer to a suitable alarm or automatic shutdown
facility.
Failure to observe this warning could cause a potentially hazardous situation which, if not
avoided, may cause death or personal injury.
2.4.1

Testing power input cables and circuit breakers

The power input cables and circuit breakers can be tested as follows.
NOTICE
This section is in accordance with IEC 60079-0: 2011 Clause 30. This section must not be
changed, amended, or removed.
Procedure
1.
2.
28
Set the main circuit breaker to ON.
Check that the display controller lights up.
The instrument begins to power up.
Section
3.3.
CT5800