2.2

Mounting

This procedure requires a minimum of two people to safely move and mount the analyzer.
Outline dimensional drawings of the analyzer are shown in Detailed System Specification.
NOTICE
In accordance with IEC 60079-0: 2011 Clause 30. The positions of the glands for the three
electrical conduits are shown in
repositioned without reference to Emerson.
WARNING!
HEAVY ITEM
In accordance with IEC 60079-0: 2011 Clause 30.
Make sure that the wall the unit is mounted on is solid, stable, and of a suitable material to
hold the weight of the analyzer.
Handle the analyzer with caution during unpacking, installing, maintaining, and transporting
to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 80 kg (176.37 lb). Emerson
and lifted by a minimum of two people. Wear suitable protective gloves and protective
footwear.
Failure to properly handle the analyzer may cause injury to personnel.
Make sure that the wall fixing points are capable of supporting a load of 80 kg (176.37 lb.)
each; this includes a x2 safety factor. All mounting points are 15 mm diameter holes.
Procedure
1.
2.
Quick Start Guide
Figure
Use safety approved equipment to lift the analyzer from the stable platform.
One person guides the unit into position as the other person carefully operates the
lifting equipment.
Mount the analyzer using the factory fitted and pre-drilled wall mounting brackets
(Figure
2-1) using four M8 x 1.25 (5/16 - 18 UNC) A2/A4 grade 70 bolts.
2-4. These positions must not be redesigned or
recommends that the analyzer is only moved
Install
11