sample line temperature is 338 °F (170 °C). During this time, the eductor air
solenoid will remain closed, so no sample is pulled through the transmitter. When
the transmitter reaches operating temperature, the solenoid will energize,
eductor air will begin to flow, and the transmitter will begin normal operation.
3.3 

Setting test gas values

3.3.1 
Set test gas values with HART
Procedure
1. Use the Field Communicator software to access the HART
2. From the DETAILED SETUP menu, select CAL SETUP.
3. From the DETAILED SETUP menu, select O2 CALIB PARAMS.
4. From the CAL SETUP menu, select O2 CAL PARAMS or COe CAL PARAMS.
5. From O2 CAL PARAMS, select O2 HIGH GAS. Enter the percent O
6. From O2 CAL PARAMS, select O2 LOW GAS. Enter the percent O
7. From the DETAILED SETUP menu, select COe CALIB PARAMS.
8. From COe CAL PARAMS, select COe Test Gas. Enter the CO concentration
3.3.2 
Set test gas values with F
Communicator
Procedure
1. Use the Field Communicator software to access the F
2. From the TRANSDUCER menu, select O
3. From the O
4. From the O
5. From O
6. From the TRANSDUCER menu, select COe CAL SETUP.
7. From COe CAL SETUP, select COe Test Gas. Enter the CO concentration
Rosemount OCX8800C
the high O
test gas.
2
the low O
test gas.
2
(ppm) used for COe test gas.
menu.
CAL menu, select O
2
CAL SETUP menu, select O
2
used for the high test gas.
CAL SETUP, select O
2
low test gas
(ppm) used for the COe test gas.
Configuration and start-up
®
OUNDATION
CAL.
2
CAL SETUP.
2
HIGH GAS. Enter the percent O
2
LOW GAS. Enter the percent O
2
December 2022
menu.
®
used for
2
used for
2
Fieldbus
Fieldbus
OUNDATION
2
used for the
2
55