Plan
WARNING!
EXPLOSION HAZARD
Installing the instrument requires opening the enclosure and working at the open instrument.
This is permitted only when both no hazardous atmosphere is present and the instrument and
connected external circuitry are de-energized.
Depending on the local regulations, this may require a competent hot work supervisor to issue
a hot work permit.
CAUTION!
SHOCK AND VIBRATION
The Rosemount CT5800 contains sensitive electronic equipment. It must not be subjected to
any shock and or vibration.
Damage to the analyzer may result from a failure to follow this caution.
Procedure
1.
2.
3.
4.
5.
6.
1.2

Installing and connecting the unit

The following notices should be followed to ensure compliance with the low voltage
directive (Europe) and other applicable regulations.
2
On receipt of goods, look for any visible damage to the analyzer and verify that all
items noted to be shipped were received. Record on the goods receipt note any
damage or missing items, noting both the item(s) and the quantity missing.
Visually inspect the exterior of the analyzer for signs of damage, corrosion, gas
leaks, or signs of previously overheating.
Report anything found to the maintenance organization.
Use safety approved and tested lifting equipment to remove the analyzer from the
shipping container.
Place the analyzer on a solid, level surface and prepare to wall mount the analyzer.
Make sure that the analyzer is stored in its protective plastic cover until installation.
Suitable grounding should be made at all connectors provided for this purpose.
All safety covers and grounding connections must be properly reinstated after
maintenance work or troubleshooting.
A fuse shall be provided at the installation site which will completely disconnect the
unit in case of failure. Installing an isolating switch may also be beneficial. In either
case, these components must be constructed to conform to recognized norms.