Connecting the signal cables

Connect the signal cables through conduit outlets 2 and 3 as shown in the figures below.
NOTICE
This section is in accordance with IEC 60079-0: 2011 Clause 30. This section must not be
changed, amended, or removed.
The signal cables are connected to the system through conduit outlets 2 and 3 as shown in
Figure 2-7
cables. Customer supplied conduit and cables must be terminated in the conduit outlets in
accordance with local electrical codes.
Figure 2-7:
A.
B.
C.
D.
Figure 2-8:
Quick Start Guide
and
Figure
2-9. All signal cables are to be minimum 20 AWG tri-rated switchgear
Cable Glands, Conduits, and Gas Line Connections
Breather
Gas connection out (flame arrestor) (M18 x 1.5)
Gas connection in (flame arrestor) (M18 x 1.5)
Conduit / cable entry apertures (M20 x 1.5)
Signal connections
Install
21