Safety Information
All authorized users, including installation, operation, and maintenance personnel, must observe the following safety precautions
and warnings.
This article is in accordance with IEC 60079-0: 2011 Clause 30.
This article must not be changed, amended, or removed.
DANGER!
FLAMMABLE SUBSTANCES
This article is in accordance with IEC 60079-0: 2011 Clause 30.
Internal parts of the analyzer may reach temperatures of 65 °C (150 °F) and may present an ignition source.
Exercise care when using oil, paint, cleaning rags, and other flammable substances near the analyzer. A fire may result if this
precaution is not observed. Always assume that the interior of a analyzer is hot unless it has been switched off and allowed to cool
down.
Failure to observe this warning will cause death, personal injury, and/or damage to persons and/or property.
DANGER!
ELECTRIC SHOCK
This article is in accordance with IEC 60079-0: 2011 Clause 30.
The analyzer operates using mains voltage that is dangerous to life. Make sure that the circuit breakers are set to OFF and tagged off
before opening the front cover.
DANGER!
FAILURE TO LOCK-OUT GAS HANDLING SYSTEM MAY CAUSE DEATH.
This article is in accordance with IEC 60079-0: 2011 Clause 30.
Always lock out the gas handling system when shutting down the analyzer. Unauthorized operation of the gas handling system
when maintenance is being performed on the analyzer or its associated pipes/hoses may result in highly flammable gas being
released, causing fire or explosion.
DANGER!
FAILURE TO VENT SAMPLE GAS MAY CAUSE DEATH.
This article is in accordance with IEC 60079-0: 2011 Clause 30.
The sample gas in the system must be vented to prevent fire or explosion during maintenance and to prevent damage to the
analyzer during startup.
The sample gas in the pipes leading to the analyzer must be purged to prevent hazards to personnel during maintenance. Purging
the sample gas must be done in accordance with the safe working procedures for the site.
Allow the analyzer and system for returning the sample gas to run for five minutes to allow any sample gas in the system to be
returned to the exhaust.
WARNING!
EXPLOSION HAZARD
The unit described in this manual may not be used in explosive atmospheres without additional safety measures.