Temperature Sensor And Cell Heater; Pressure Sensor - Emerson Rosemount CT5800 Quick Start Manual

Continuous gas analyzer
Hide thumbs Also See for Rosemount CT5800:
Table of Contents

Advertisement

2.4.2

Temperature sensor and cell heater

The sample cell is controlled to operate at a pre-set temperature. This involves a heater
and a temperature sensor.
NOTICE
This section is in accordance with IEC 60079-0: 2011 Clause 30. This section must not be
changed, amended, or removed.
Procedure
1.
2.
2.4.3

Pressure sensor

The pressure sensor monitors the pressure in the analysis cell.
NOTICE
This section is in accordance with IEC 60079-0: 2011 Clause 30. This section must not be
changed, amended, or removed.
Procedure
1.
2.
Quick Start Guide
Check that a temperature reading is displayed against Temperature on the display
controller as described in
Leave the analyzer operating and confirm that the temperature rises until the
analysis cell reaches the pre-set operating temperature around 50 °C (122 °F).
This takes approximately 90 minutes.
Check that a pressure reading is displayed under Pressure on the display controller.
The reading is approximately 760 Torr at atmospheric pressure.
NOTICE
A Torr is a non SI unit of pressure, defined as 1/760 of standard atmospheric pressure,
and is equal to the fluid pressure of 1 mm of mercury.
If desired, cap off the gas inlet and use an external pump to evacuate the cell.
Make sure the pressure drops as expected.