Table 2-1:
Electrical power requirements
Electrical supply
Instrumentation supply
voltage
Purge supply voltage
For the electrical power wiring, use 16 AWG stranded, 3 conductor copper or tin plated
copper power wire, rated for at least 250 Vac, of the required length.
Cables must be rated for operation in ambient temperatures > 80 °C (176 °F).
Cables must be terminated in accordance with local electrical codes.
A switching system is not supplied with this equipment. You must supply a suitably rated
switch or circuit breaker to be included with this installation. Check the installation of the
switch for conformity in accordance with national / local regulations and standards by
inspection.
The switch or circuit breaker must be suitably located, easily reached, and identified as the
disconnection device for the analyzer.
Figure 2-4:
A.
B.
C.
D.
The customer supplied circuit breaker must be in accordance with suitably certified
hazardous area Ex protection concepts. The main power isolator controls the application
of electrical power to the analyzer.
Quick Start Guide
Power consumption
500 W (peak)
300 W (usual)
10 W
CT5800 Bottom view - Gas connectors and cable glands
Gas connections (flame arrestor) (M18 x 1.5)
Conduit / cable entry apertures (M20 x 1.5)
Wall fixing line
Earth (ground) bond
Voltage
110 to 240 Vac, 50/60 Hz
± 10%
110 to 240 Vac, 50/60 Hz
± 10%
Install
Fuse
3.15 A internal fuses F1 and
F2. See
Section
2.3.4.
1 A (located in secondary
circuit breaker)