Install
In accordance with IEC 60079-0:2011 Clause 30, install the analyzer in a suitable position
with shading to protect it from the elements. The displays and control panel on the front of
the housing must not be exposed to direct UV light sources or direct sunlight.
WARNING!
POTENTIAL ELECTROSTATIC CHARGING HAZARD
This article is in accordance with IEC 60079-0: 2011 Clause 30.
The keypad is non-conducting and may generate an ignition capable level of electrostatic
charges under certain extreme conditions. Ensure that the equipment is not installed in a
location where it may be subjected to external conditions (such as high pressure steam) which
might cause a build-up of electrostatic charges on non-conducting surfaces. Additionally, only
clean the equipment with a damp cloth.
The analyzer has a T3 temperature classification which specifies the maximum surface
temperature of the instrument, under a fault condition, is 200 °C (392 °F). You must ensure
that no combustible gas concentrations are present, whether on a continual or occasional
basis, which have an ignition temperature below the T classification of the unit.
WARNING!
FIRE AND EXPLOSION
Do not open the Ex d enclosure of the analyzer unless the atmosphere in the area is known to
be below the ignitable concentration of combustible gases or materials.
Failure to observe this precaution will cause death, personal injury, and/or damage to persons.
In accordance with IEC 60079-0: 2011 Clause 30, the analyzer must be installed in a
suitable position with shading to protect it from the elements. The displays and control
panel on the front of the housing must not be exposed to direct UV light sources or direct
sunlight.
A fuse must be provided at the installation site which will completely disconnect the
analyzer in case of failure. Installing an isolating switch may also be beneficial. In either
case, these components must be constructed to conform to recognized norms.
In accordance with IEC 60664-1, the analyzer must be installed in an area of not more than
Pollution Degree 2.
Provide sufficient space around the analyzer to allow maintenance and servicing of this
unit.