Measurement Complete signal:
/O_INDEX: This signal will go the low level when analog measurement is
!
Compare Output Signals:
/O_BIN_1- /O_BIN_13 : The GO judgment is active according to the
/O_P_HI: Main parameter failure: When the main parameter is greater than
Max, it will go the low level till T4 (see Handler Interface
Timing).
/O_P_LO: Main parameter failure: When the main parameter is lower than
Min, it will go the low level till T4(see Handler Interface
Timing).
/O_P_OVER: It will go the low level when main parameter is greater than
/O_S_REJ,/O_S_OVER: It will go the low level when secondary is greater
!
Judgment Complete Signal
/O_EOM: This signal becomes active (low level) when the measurement
and comparator judgment are completed, and all handler
interface output are stable. It will go the high level when next
trigger Signal (/I_E_TRIG) becomes active (low edge).
completed and will go the high level when next trigger is
active. The handler can then remove the DUT from the fixture
and replace it with another DUT during the time when
/O_INDEX is at low level.
measured value judgment results (low level),
for example, if the comparator assigns a
component to BIN1, the /O_BIN_1 line will go
the low level till T4 (see Handler Interface
Timing). /O_BIN_2 - /O_BIN_13 are all
inactive (high level).
Max or is lower than Min.
than D_Max (if MODE=C/D) or is lower than
D_Min.