The other BIN setting:
!
When the BIN 1 setting is finished, press F4 key to continue the BIN 2
setting until all the Tot BIN setting is completed (BIN1~BIN13 the most),
please refer to Figure 8-22.
SORT BY C
C_Nom.Val: 15.000pF
Max:16.500pF+:10.0%
Min:13.500pF-:10.0%
D_Max:1.000
D_Min:.0010
TEST RESULT Display Summing-up:
!
Press F1/F2 key to select the white inverted characters "BIN.SUM" (Figure
8-23).
!
Press F3 key to see the test result.
Sorting judging rule: To check whether the secondary parameter is within the
test range, if not, just display SREJ instead of judging the primary
parameter, if yes, continue to proceed primary parameter sorting.
!
Press F4 key to delete test result (Figure 8-24).
!
Press F3 orF4 key to select delete or non-delete test result (Figure8-25).
!
Press F2 key to see next page of test result (Figure 8-26, 8-27).
BIN1
TOt_Bin:13
Figure 8-22
BIN
CLR
SORT
D
Press F4 key
NEXT
BIN
EXIT
58