Gas Requirements and Configurations

Agilent 4100 MP-AES Gas Configurations

Your Agilent 4100 MP-AES may come with one or more Agilent
accessories. Some of the accessories also require gas supplies. This
section is designed to assist you with planning your gas requirements
according to the system and accessories ordered so that at the time
of installation you have the necessary gas supplies, and they meet the
required specifications.
Shown below is a summary table detailing the various gas
requirements for the Agilent 4100 MP-AES and accessories. Table 14
details nitrogen and argon gases, and the compressed air
requirement, for the POP gas.
Table 14. Nitrogen and argon specifications
Intended use
Purity
Oxygen
Water
Required flow
(SLPM)
Required
pressure (kPa)
Table 15 details the compressed air requirements for the Agilent
4100 MP-AES and accessories.
48
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Nitrogen
Plasma support gases
>99.5%
< 0.5%
< 4ppm
25
450–600
Nitrogen
Argon
Purged mono
Plasma ignition gas
>99.95%
99%
< 0.05%
n/a
< 4ppm
n/a
10
1.5
450–600
200–440