Laboratory Facilities; Exhaust System - Agilent Technologies 4100 Site Preparation Manual

Microwave plasma atomic emission spectrometer
Hide thumbs Also See for 4100:
Table of Contents

Advertisement

This section includes information on the laboratory facilities
required to support the operation of the instrument.

Exhaust System

The plasma operates at a temperature of approximately 6,000 °C,
(11,000 °F) with up to 1,000 Watts of power (6,480 kilojoules per
hour). The instrument exhaust fumes can be noxious or corrosive.
The Agilent 4100 MP-AES instrument chimney must be connected to
an exhaust system. The exhaust system must include an exhaust fan,
be ducted to an external vent and provide a minimum flow of
3 m
The exhaust system installation must comply with any rules and/or
regulations that may be imposed by the local authorities responsible
for control of facilities and fixtures in the work place.
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide

5. Laboratory Facilities

3
/min (106 cfm) at the intake point.