Electrical Requirements; Weights And Dimensions - Agilent Technologies 4100 Site Preparation Manual

Microwave plasma atomic emission spectrometer
Hide thumbs Also See for 4100:
Table of Contents

Advertisement

Accessories and Options

Electrical requirements

All power supplies should be single phase AC, three-wire system
(active, neutral, ground or two active and ground) and should be
terminated at an appropriate connection receptacle that is within
reach of the power cable.
The installation of electrical power supplies must comply with the
rules and/or regulations imposed by the local authorities responsible
for the use of electrical energy in the workplace.
The Agilent 4107 Nitrogen Gas Generator is supplied with a 2 meter
long (6 ft 6 in), mains supply cable. The cable is rated to 15 A/250 V
and is terminated as listed in Table 12.
Table 11. Nitrogen generator electrical specifications
System unit
Nitrogen generator
Plug pack
Table 12. System electrical specifications
Power supply phase
Australia
USA
Europe

Weights and dimensions

See Page 26 for details.
44
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Required supply voltage
24 V DC ±5%
100–240 V AC ±10%
Frequency 47–63 Hz
Single
Equipment end
IEC-60320-C13
IEC-60320-C13
IEC-60320-C13
Nominal rating
50 W
60 W
(Nominal)
Supply end
AS/NZS-3112 10 A 3 pins
NEMA 5-15P
CEE 7/7