This document describes the installation of the field kits for the manual
injection ports (septum nut and luerlock) at the front of the Agilent 490
Micro GC.
When the Agilent 490 Micro GC is equipped with a standard unheated front inlet, an
N O T E
additional field upgrade kit for the sample inlet box is required. The part number for a dual
cabinet Agilent Micro GC is CP490202, for the quad cabinet Agilent Micro GC use part
number CP190203.
Agilent 490 Micro GC
Manual Injection Port Field Kit
Installation Instructions
Agilent Technologies

Advertisement

loading

Summary of Contents for Agilent Technologies 490 Micro GC

  • Page 1 (septum nut and luerlock) at the front of the Agilent 490 Micro GC. When the Agilent 490 Micro GC is equipped with a standard unheated front inlet, an N O T E additional field upgrade kit for the sample inlet box is required. The part number for a dual cabinet Agilent Micro GC is CP490202, for the quad cabinet Agilent Micro GC use part number CP190203.
  • Page 2 Agilent 490 Micro GC Table 1 Parts list for the four kit types Kit types available  CP490204 CP490205 CP490206 CP490207 Description and Part number Septum Luerlock Septum + Rear Luerlock + Rear Bracket with pressure relief valve CP742849/CP742987 Bracket with pressure relief valve + ball-valve...
  • Page 3: Mounting The Sample-In At The Front Of The Agilent 490 Micro Gc

    Agilent 490 Micro GC Mounting the Sample-in at the Front of the Agilent 490 Micro GC Remove the side and front panels of the Agilent 490 Micro GC. Make a hole in the preferred place on the front panel next to Sample 1. See...
  • Page 4 Agilent 490 Micro GC If using the Injection-Port CP490204 or CP490206 use the parts listed in Figure 1. CP742983 2. CP72472 3. CP86757 4. CP739952 Figure 3 Sample bracket parts for CP490204 or CP490206 If using the Injection-Port CP490205 or CP490207 use the parts listed in Figure 1.
  • Page 5 Sample bracket parts for CP490205 or CP490207 The diagrams below show the differences for the different field kits (Figure 6 Figure 7). Plumb the tubing according to these diagrams and Figure Figure 6 Diagram of unheated sample-in-Agilent 490 Micro GC for CP490204 or CP490205 Installation Instructions...
  • Page 6 Agilent 490 Micro GC Figure 7 Diagram of unheated sample-in-Agilent 490 Micro GC for CP490206 or CP490207 Figure 8 Diagram of unheated sample-in-Agilent 490 Micro GC Installation Instructions...
  • Page 7 Agilent 490 Micro GC Make an 18 mm incision in the upper insulation plate as seen in Figure 9 and install it in the Sample-Inlet-Box, see Figure 10 Figure 18 mm Figure 9 Upper insulation plate Figure 10 Sample-Inlet-Box without insulation Figure 11 Sample-Inlet-Box with insulation Reinstall the side panel.
  • Page 8: Agilent Technologies

    © Agilent Technologies, Inc. 2017 Printed in USA and China Second edition, August 2017 Agilent Technologies, Inc.