Laboratory Facilities
The exhaust fan should be located at least 2 meters (6 ft 6 in) away
from the top of the instrument chimney. The fan control switch and
running indicator lamp should be located in a position where the
instrument operator can view the indicator and access the control
switch.
Ducting must be corrosion resistant, fireproof and should be kept
clear of fire alarms, sprinkler heads and other heat sensitive devices.
The external vent must be fitted with a back draft damper and the
outlet location must be clear of doors, windows and heater or air-
conditioning units.

Electrical Power Supplies

The installation of electrical power supplies must comply with the
rules and/or regulations imposed by the local authorities responsible
for the use of electrical energy in the workplace.
All power supplies for the Agilent 4100 MP-AES instrument and its
accessories should be single phase, AC, 3 wire systems (active,
neutral, ground or two active and ground). Each connection should
be terminated at an appropriate receptacle within reach of each
assembly's power cable. Use of power boards or extension cables is
not recommended.
Avoid using power supplies from a source that may be subject to
electrical interference from other services (such as large electric
motors, elevators, welders and air conditioning units).
The Agilent 4100 MP-AES instrument is supplied with a 2 meter long
(6 ft 6 in), mains supply cable. The power cable is rated to 10 A/250 V
and terminated as indicated in Table 5.
30
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide