Electrical requirements
Table 4. System electrical specifications
System unit
4100 MP-AES
SPS 3 autosampler
Personal computer
(Typical)
Printer
(Typical)
Instrument power connections
Table 5. Instrument power connections
Power supply phase
Cable rating
Australia
USA
Europe
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Required supply voltage
200–240 V AC ±10%
Frequency 50–60 Hz ±1 Hz
96–264 V AC
50–60 ±1 Hz
115, 120, 220, 240 V AC, 50/60 Hz
115, 120, 220, 240 V AC, 50/60 Hz
Single
10 A/250 V AC
Equipment end
IEC-60320-C13
IEC-60320-C13
IEC-60320-C13
Laboratory Facilities
Nominal rating
2.04 kVA
~220 VA
300 VA
85 VA
Supply end
AS/NZS-3112 10 A 3 pins
NEMA 6-15P
CEE 7/7
31