Gas Requirements and Configurations
Table 13. Air compressor requirements
Equipment
Nitrogen Generator
POP, EGCM,
Monochromator Air
Purge
The nitrogen generator will be permanently damaged by oil vapor or aerosol from
CAUTION
a compressor. The damage will not be repairable and will require replacement of
the product
Oil-free compressors may still deliver oil (hydrocarbons) to the nitrogen generator
CAUTION
if the inlet air is contaminated. It is essential that clean air, free of hydrocarbons
and acid, is supplied to the inlet of the compressor. A three-stage filter is used
between the compressor and nitrogen generator to protect the nitrogen
generator from any remaining contamination.
Even when an oil-free compressor is used the filter unit incorporated in the
CAUTION
nitrogen generator must be used.
46
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Quality
Pressure
Specification
ISO 8573-1:2010
620–750 kPa (90–109 psi)
CLASS 8.4.3
ISO 8573-1:2010
450–600 kPa (58–87 psi)
CLASS 1.4.3