Figure 6. Agilent 4100 MP-AES, N
supplied from Onboard Argon Supply, POP gas supplied from the air compressor, EGCM configured
to purge the monochromator with air from the air compressor and to provide air for analysis of
organics.
Oil Free Air
Compressor
and Dryer
Total Compressed Air = 115+10+25+1.5 = 151.5 SLPM
Total Nitrogen (99.5%) = 0 SLPM
Total Nitrogen (99.95%) = 0 SLPM
Total Argon = 1.5 SLPM
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
plasma gases supplied from nitrogen generator, argon gas
2
Agilent
Nitrogen
Generator
Filter/ Regulator
Gas Requirements and Configurations
MP-AES
MP-AES Plasma
Gas Box
EGCM