Accessories and Options
Gas requirements
All gas supply installations must comply with rules and/or
regulations that are imposed by the local authorities responsible for
the use of compressed gas in the workplace.
Table 8 Monochromator Optics Air Purge
Quality
Permissible pressure range
Recommended pressure
Required flow rates
Failure to use the specified air quality will significantly may result in
NOTE
malfunction or damage to the 4100 MP-AES.
Sample Preparation System (SPS 3)
The Agilent Sample Preparation System 3 (SPS 3) is compatible with
a wide range of commercially available low-cost autoclavable sample
racks.
Sample contamination from airborne particles is eliminated and
corrosive or toxic fumes are removed during sampling with the
optional environmental enclosure.
40
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Air
ISO 8573-1:2010 CLASS 1.4.3
450–600 kPa (58–87 psi)
600 kPa (87 psi)
10 SLPM