Table 15. Compressed air specification for the nitrogen generator and accessories
Intended use
POP gas
Quality
ISO 8573-
1:2010 CLASS
1.4.3
Required
25
flow (SLPM)
Required
450–600
pressure
(kPa)
Max output
n/a
flow (SLPM)
Output
n/a
pressure
(kPa)
The following pages detail some the possible 4100 MP-AES and
accessory configurations along with their associated gas
requirements. Use these as an example for calculating/determining
your gas requirements.
Agilent 4100 Microwave Plasma Atomic Emission Spectrometer Site Preparation Guide
Nitrogen
Monochromator
generator
nitrogen
plasma gas
generator purge
ISO 8573-
ISO 8573-1:2010
1:2010 CLASS
CLASS 8.4.3
8.4.3
115
115
620–750
620–750
25
10
450–600
450–600
Gas Requirements and Configurations
Organics
ISO 8573-
1:2010
CLASS 1.4.3
1.5
450–600
n/a
n/a
Monochromator
air purge
ISO 8573-1:2010
CLASS 1.4.3
10
450–600
n/a
n/a