Figure 5. Agilent 4100 MP-AES, N
supplied from Onboard Argon Supply, POP gas supplied from the air compressor, EGCM configured
to purge the monochromator with N
compressor for analysis of organics.
Oil Free Air
Compressor
and Dryer
Filter/ Regulator
Total Compressed Air = 115+115+25+1.5 = 256.5 SLPM
Total Nitrogen (99.5%) = 0 SLPM
Total Nitrogen (99.95%) = 0 SLPM
Total Argon = 1.5 SLPM