Laboratory Supply Requirements

The following items are commonly used with the 7700 ICP- MS system. Refer to
Table 14
A set of clean hand tools dedicated for maintenance of ultra- clean parts such as
the ion lens, interface, etc.
Reagent- grade acetone, methanol, or isopropanol for cleaning. Proper storage,
handling, and disposal of these chemicals is required for personal and
environmental safety.
Laboratory glassware detergent and deionized water for general cleaning of
system components.
Chemical solvents should be considered hazardous and must be handled with care.
WA R N I N G
Contact your chemical supplier for additional solvent handling and safety
information.
An auxiliary work space and fume hood are needed for maintenance procedures
such as interface, ion lens, etc. cleaning.
Table 14 Recommended Tools, Equipment and Supplies
Item
Ball Driver
Beakers
Sample Bottles
Cloths
Cotton Applicators
Gloves
Gloves
Magnifier
Ultrasonic Bath
Pliers
Screwdrivers
Wrenches
Volt/Ohm Meter
G3280-90010 (Rev.B) Agilent 7700 Series ICP-MS Site Preparation Guide
and you must order necessary supplies.
Description
Metric, Hex set
50mL, 250mL, 600mL
Polypropylene
Clean Lint Free
Lint Free
Chemical Resistant
for Inspection of Interface Cones
at least 1 Quart Capacity
Needle Nose
Torx, Phillips & Flat head
Various Metric
High Impedance > 10M ohm
29