Perform A System Check; Pause Or Stop A Run - illumina MiSeq System Manual

Hide thumbs Also See for MiSeq:
Table of Contents

Advertisement

MiSeq System Guide
When the files finish bundling, the Bundle Logs tab reopens.
8
Send the zipped bundle to Illumina Technical Support.
Figure 23 Bundle Logs Tab

Perform a System Check

The System Check screen is typically used to connect with an Illumina Technical Support representative
during a Live Help session. Use of this feature is not required during normal operation or for instrument
maintenance.
Some system checks can be performed before contacting Illumina Technical Support, such as the Volume
Test. A volume test checks the health of the fluidics system by estimating the flow volume as bubbles pass by
the sensors. For more information, see
1
From the Home screen, select Manage Instrument.
2
Select System Check.
3
Do one of the following:
Select the individual tests you want to perform.
u
Choose Select All to perform all tests.
u
4
Select Next.
When complete, the test results appear on the screen.
5
[Optional]
Select Show Details to see a summary of the results on the software interface.
6
[Optional]
Select Export Results to export the results in a *.csv file format to a USB drive.
7
Select Done.

Pause or Stop a Run

The MiSeq is designed to complete a run from beginning to end without user intervention. However, it is
possible to pause a run or stop a run from the Sequencing screen.
Document # 15027617 v04Material # 20024228
For Research Use Only. Not for use in diagnostic procedures.
Perform a Volume Test on page
45.
42

Advertisement

Table of Contents
loading

Table of Contents