11 Remove the waste bottle and discard the contents appropriately. Return the waste bottle to the reagent
compartment.
WARNING
This set of reagents contains potentially hazardous chemicals. Personal injury can occur through inhalation,
ingestion, skin contact, and eye contact. Wear protective equipment, including eye protection, gloves, and
laboratory coat appropriate for risk of exposure. Handle used reagents as chemical waste and discard in
accordance with applicable regional, national, and local laws and regulations. For additional environmental,
health, and safety information, see the SDS at support.illumina.com/sds.html.
12 Slowly lower the sipper handle, making sure that the sippers lower into the wash bottle and waste bottle.
13 Close the reagent compartment door.
14 Select Next.
When the wash is complete, leave the used flow cell, wash tray, and wash bottle containing the remaining
wash solution on the instrument.
NOTE
The sippers remain in the down position. Leave the unused wash solution in the wash tray and wash bottle
to prevent the sippers from drying out and air from entering the system.