0 - Preface Manufacturer 0 Preface For Your Safety Before performing any work on or with the Infinium LiHa and Infinium RoMa, first read the Operating Manual - Document # 1000000110155 v00 carefully, in particular chapter 2 “Safety”. 0.1 Manufacturer...
CE Conformity 0.3 CE Conformity Declaration of The Infinium LiHa and Infinium RoMa is designed and built in compliance with the basic safety and health requirements of applicable EC Directives. With the Conformity declaration of conformity the manufacturer declares conformity with the provisions of the Directives.
Operation of this equipment in a residential area is likely to cause harmful interference in which case users will be required to correct the interference at their own expense. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 0 - 3...
Page 6
0 - Preface FCC Rules 0 - 4 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Target Group This manual is intended for everyone who wants to learn about the safe operation of the Infinium LiHa and Infinium RoMa and how to maintain its perfect working condition. In particular, laboratory personnel and operators are addressed. Laboratory personnel operating Infinium LiHa and Infinium RoMa instruments require also thorough knowledge of applications, instrument functions and software programs as well as all applicable safety rules and regulations.
Instruments Any individual or separate operating manuals for optional equipment according to your order configuration can be applicable. For information about the liquids to use with the Infinium LiHa and Infinium RoMa, refer to section 3.5 “Chemical Resistance”, 3-17.
Pressure based liquid level detection Pressure monitored pipetting Polyoxymethylene Polypropylene PosID Positive identification option, barcode scanner Polystyrene PTFE Polytetrafluoroethylene Polyvinylchloride PVDF Polyvinylidenefluoride RoMa Robotic manipulator arm Radio frequency Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 1 - 3...
Page 14
1 - About This Manual Abbreviations Sensored pump option Universal serial bus Uninterruptable power supply World Health Organization 1 - 4 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
2 - Safety Safety Message Conventions 2 Safety This chapter describes the safety concept of Infinium LiHa and Infinium RoMa, provides general rules of correct behavior, and warnings concerning hazards associated with the use of the Infinium LiHa and Infinium RoMa.
Page 16
2 - Safety Safety Message Conventions Hot surface hazard Wear protective gloves Read this Disturbance of functions by electromagnetic RF waves. Do not use a cellular phone. 2 - 2 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Nevertheless, risks to users, property and the environment can arise if the Infinium LiHa and Infinium RoMa is used without due care and attention. The safety of all users and personnel depends on the strict observation of these safety instructions and awareness of the safety-related warnings provided in this manual.
Page 18
Infinium LiHa and Infinium RoMa. The same applies to waste disposal. – Always be aware of possible hazards associated with these substances. 2 - 4 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
For more information go to www.P65Warnings.ca.gov/product. 2.3 Operating Company The operating company must ensure that the Infinium LiHa and Infinium RoMa, in particular the safety features, function properly and that all the personnel in contact with the instrument are adequately trained.
User Qualification 2.4 User Qualification The laboratory personnel must be fully qualified and trained to operate the Infinium LiHa and Infinium RoMa. The work described in this Operating Manual must only be performed by authorized personnel with the qualifications prescribed below.
Door Locks During operation the front safety panel is locked by means of two door locks. The safety concept of the Infinium LiHa and Infinium RoMa assumes that the front safety panel is always closed when the instrument is running.
Page 22
Side safety panel Cutout for continuous loading General Removal of The protective and safety devices installed on the Infinium LiHa and Infinium RoMa must not be removed or disabled during operation. Safety Elements In the event such elements were removed, e.g. for maintenance work, operation may only be resumed when all protective and safety devices have been completely installed and checked.
2.6 Product Safety Signs Where are Infinium LiHa and Infinium RoMa Instrument Safety Notices The figure shows the safety notices that are attached to the Infinium LiHa and Attached? Infinium RoMa instrument. It also shows their locations: Fig. 2-2 Safety notices attached to the product The following table explains the significance of the notices: Tab.
Class 2 Laser Product pursuant to IEC 60825-1:2007: "Complies with 21 CFR 1040.10 except for deviations pursuant to Laser Notice No. 50, dated June 24, 2007" 2 - 10 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 25
On barcode scanner the scanner head assembly by hand which could head, see Fig. 2-5, 2-11 damage motor and head assembly. a) According to IEC/EN 60825-1 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 2 - 11...
Page 26
The only possible health hazard is in exposing the eyes to the laser beam. Do not stare into the beam. 2 - 12 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Repair Order) in case the owner of the instrument has no template for such a declaration at hand. Contact the Illumina helpdesk for further information. Note: Illumina reserves the right to refuse any instrument or a part of it, or will charge an extra fee, if the decontamination is not declared sufficiently.
Page 28
2 - Safety Decontamination Declaration 2 - 14 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
3 - Technical Data Introduction 3 Technical Data Purpose of This This chapter introduces the reader to the Infinium LiHa and Infinium RoMa and its main components. It contains technical data, requirements and performance data. Chapter 3.1 Introduction What is The Infinium LiHa and Infinium RoMa is a precision instrument designed for automating specific steps in the Illumina Infinium Array Workflow.
The identification data is also printed on the serial number label (B). This label can be viewed from the front side of the instrument. It is attached below the diluters. Label 3 - 2 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
The figure shows the dimensions of the instrument: Overall Dimensions 780 mm (30.7 1224 mm (48.2 *maximum height for objects under the x rail Fig. 3-2 Instrument overall dimensions Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 3 - 3...
(100 - 120 VAC) must not exceed 1000 VA in order to keep the input current below the fused values. 3 - 4 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
When the status lamp light is red, an acoustic alarm sounds (status lamp functions: Refer to 6.2 “Operating Modes”, 6-3). The illuminated area is 540 x 18 mm (21.26 x 0.7 in.). Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 3 - 5...
Conditions Transport temperature -20°C to 60°C (-4°F to 140°F) for maximum 24 hours Transport humidity 20% to 80% relative (non condensing) for maximum 24 hours 3 - 6 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
The instrument is connected via USB to the control PC. Tab. 3-6 USB Interface Specification (Instrument) Interface Voltage Current USB 2.0 4.4 V to 5.25. V max. 500 mA Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 3 - 7...
-the art by the kit manufacturer or operator of the system. For sample preparation refer to cross references above. 3 - 8 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Y-axis 8-tip LiHa 373 mm (14.7 in.) Z-axis 210 mm (8.27 in.) a) at 9 mm spacing b) each channel individually, no tips installed Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 3 - 9...
Page 38
The equidistant movement of sampling tips in Y direction is: Movement from 9 mm ± 0.4 mm to 38 mm ± 1 mm 3 - 10 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 39
PCTFE (Kel-F) Syringes Borosilicate glass Syringes, seals PTFE Tips Stainless steel, PTFE Liquid containers HD-PE a) Coating Also refer to section 3.5 “Chemical Resistance”, 3-17. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 3 - 11...
270° (left or right oriented) ATTENTION Improper transport of labware (microplates, etc.) Use only labware that is rigid enough not to be deformed by the gripper force. 3 - 12 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Container ID barcode, 100 ml trough on carrier 100 mm/s 99.8% a) Barcode scanner b) “No Tube” sensor, glass or plastic tube, filled or empty, with or without barcode Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 3 - 13...
Page 42
Also used for standard carrier ID barcodes b) The application SW may restrict the use of barcode types. See section “Permissible Barcode Types” below. 3 - 14 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 43
To avoid misreadings, the quality of the barcode labels must be graded A, B or C according to ANSI X3.182 and DIN EN 1635. Illumina recommends using grade A for best reading performance. A quality system in the production of barcode labels must be employed to ensure the conformity to the quality grades mentioned above.
Page 44
Note: Depending on the application software, not all barcode types readable by the PosID may be supported. Refer to your Application Software Manual to find out the permissible barcode types. 3 - 16 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 46
32 % Trichloroacetic acid 40 % a) Kel-F Legend: resistant partly resistant, use is possible with frequent replacements not resistant, unsuitable for use not determined 3 - 18 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
4 - Description of Function Introduction 4 Description of Function Purpose of This This chapter explains the basic principle of the Infinium LiHa and Infinium RoMa, shows how it is structured and gives a functional description of the assemblies. Chapter 4.1 Introduction...
4 - Description of Function Structure 4.2 Structure 4.2.1 Mechanical Structure The figure shows the main parts of the Infinium LiHa and Infinium RoMa: Fig. 4-1 Infinium LiHa and Infinium RoMa instrument overview Housing Worktable with positioning pins Diluters with syringes...
Waste tubing Syringe In case of a second liquid handling arm, each LiHa is equipped with its own parts, i.e. the two liquid systems are independent. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 4 - 3...
The liquid handling arm is part of the liquid system and is used for pipetting tasks. Fig. 4-3 The liquid handling arm, LiHa Syringes Tips Liquid handling arm 4 - 4 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 51
The LiHa arm holds electronics to detect the level of liquids in plate wells, tubes and liquid containers. For details refer to section “”, 4-14. Detection Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 4 - 5...
Robotic manipulator arm RoMa Axis for gripper movements Axis from front to back of worktable Rotational axis Vertical axis above worktable Axis from left to right of worktable 4 - 6 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Illumina or Clarity LIMS, or their own custom LIMS system. The Illumina 1D barcode scanner kit is installed on the RoMa arm for all such customers. It is mounted as per ILMN FSB 506 on the bottom of the RoMa arm using special mounting hardware and an adjustable dual bracket, then connected electrically via its cable to the existing RoMa arm cable.
The front safety panel is secured in closed position with the door locks. Panel According to the size of the Infinium LiHa and Infinium RoMa and the kind of front safety panel, one or two gas springs facilitate the opening of the panel.
Page 55
(B), which is mounted to the safety panel. A switch in the locking device monitors if the safety panel is open or closed. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 4 - 9...
This improves identification security for the containers. 4 - 10 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 57
The figure shows how vertical barcodes (e.g. on tubes or reagent troughs) are scanned. Fig. 4-10 Barcode scanner position for scanning vertical barcodes PosID body Container barcode label Barcode scanner Carrier Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 4 - 11...
Page 58
Fig. 4-12 Detectable situations in a tube rack Tubes with readable barcode Tubes without barcode (or incorrect positioned barcode) No tubes present Carrier ID barcode 4 - 12 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 59
ID with the database, the software is able to identify the carrier’s properties. Barcode Labels For detailed information on barcode types and proper positioning of barcode labels on carriers and containers refer to cross references above. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 4 - 13...
Wash station Distributor 1:4 (1:2 for 2-tip instrument) Waste tubing Interconnecting tubing Waste container 3-way valve Syringe Note: The arrows indicate the direction of flow. 4 - 14 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Standard tubing system composed of PVC/silicone/PP/ Pipetting With all tubing systems, the pipetting tubing is made of FEP, which is resistant against a wide range of liquids. Tubing Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 4 - 15...
Page 62
4 - Description of Function Liquid System 4 - 16 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
5 - Putting into Operation Installation 5 Putting into Operation Purpose of This This chapter describes how the Infinium LiHa and Infinium RoMa is installed and gives instructions on initial operation. Chapter 5.1 Installation 5.1.1 Initial Installation of the Instrument The initial installation of the instrument may be done by a qualified Illumina service person only.
Page 64
5 - Putting into Operation Startup 5 - 2 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
6 Operation Purpose of This This chapter explains the operating elements and possible operating modes. It gives instructions on how to operate the Infinium LiHa and Infinium RoMa properly Chapter and safely. For the options used by your configuration, refer to the separate operating manuals.
Page 66
Display functions and controls are available in the software packages and user interfaces on the PC. Depending on your application, refer to the relevant separate documentation. 6 - 2 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
6 - Operation Operating Modes 6.2 Operating Modes Possible The Infinium LiHa and Infinium RoMa can be run in two different operating modes: Operating Routine operation mode (operator) Modes – This is the normal operating mode, in which the application is run.
Page 68
To ensure a proper liquid flow, make sure that the tubings are not twisted or inhibited from free flow. ATTENTION Instruments are intended for indoor operation with controlled temperatures.It is important to maintain constant temperature plus air gaps. 6 - 4 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 69
Using liquids with undissolved particles could lead to clogged tips and thus result in liquid not being dispensed. Clogging can also result if the tips have not been thoroughly washed. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 6 - 5...
Page 70
6 - Operation Operating in Routine Operation Mode Applications For all applications of the Illumina instrument, the user must ensure that the requirements of each protocol are carefully observed. Attention must be given to: Sample/reagent volumes and concentrations ...
Automatically moving parts. If the safety panels are not in place, injuries (crushing, piercing) are possible. Before starting the Infinium LiHa and Infinium RoMa, make sure that the safety panel is closed. Never operate the instrument with the panel open.
Page 72
6 - Operation Operating in Routine Operation Mode To switch the Infinium LiHa and Infinium RoMa on, proceed as follows: Press the power ON/OFF switch for 0.5 seconds to switch the instrument on. Wait until the status light in the power ON/OFF switch is lit.
Improper positioning of objects on the worktable may lead to disturbances or errors in the process, e.g. misinterpretation of barcodes. Do not use free space on the worktable to deposit any objects Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 6 - 9...
Page 74
RoMa are dropped after a restart. Visually inspect the arm devices whether they still hold any objects between their grippers. Remove such objects before starting up the instrument. 6 - 10 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 75
Make sure that the correct rack is used for the carrier. If a carrier is damaged, replace it immediately. Carrier ID Each carrier ID must be unique. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 6 - 11...
Page 76
Microplates must be positioned correctly on the carrier, seating well in their holder. Make sure that the microplate does not rest on the holder rim in a slanting position. 6 - 12 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 77
Make sure that all tubes are positioned correctly in the carrier and touch the rack bottom, otherwise liquid level detection and clot detection might not work properly. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 6 - 13...
Page 78
Note: The filling level of tubes, troughs and containers must not exceed 80% to avoid spillage during PosID reading. Tab. 6-2 Minimum inner diameter for primary sample tubes Tip type Tube diameter, inside Fixed tips 7 mm 6 - 14 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 79
Separate the bypass tubing and the aspiration tube in such a way that no air bubbles can be aspirated. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 6 - 15...
Page 80
The waste tubing must not sag (back pressure). The lower end of the waste tubing must not be in the liquid (back pressure). 6 - 16 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Press the power ON/OFF switch and keep it pressed for at least 2 seconds. ATTENTION Wait until the status light in the power switch is off (for approx. 10 sec.) before switching the instrument on again. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 6 - 17...
For detailed descriptions of the maintenance tasks, refer to 7 “Preventive Maintenance and Repairs”, in this Operating Manual - Document # 1000000110155 v00. 6 - 18 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Principle Only operate the Infinium LiHa and Infinium RoMa when it is in good working condition. Strictly observe the maintenance instructions as set out in this manual. To achieve specified performance and reliability of the instrument, regularly carry out the maintenance and cleaning tasks.
Page 84
Please contact www.deconlabs.com the manufacturer Liqui-Nox Weak detergent Alconox Please contact www.alconox.com the manufacturer a) These are identical products; are hereafter called Decon / Contrad 7 - 2 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 85
PosID scanner head Alcohol laser beam output win- Arm guide, arm guide Do not use any agent roller of arms Z-rod Do not use any agent Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 7 - 3...
Page 86
Cleaning Cleaning Tissue Use a lint-free tissue together with the appropriate cleaning agent. 7 - 4 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
(depending on configuration) maintenance carried out by a Illumina authorized field service engineer (FSE) is recommended. Maintenance Note: In order to be able to track all maintenance performed on the Infinium LiHa Record and Infinium RoMa over the whole lifetime, the periodic maintenance must be recorded as follows: •...
Page 88
Tab. 7-6 Daily maintenance during the day Instrument/ Component Maintenance Task Reference Liquid system Flush prior to each application See section 7.3.1.2 “Flushing the Liquid System”, 7-12 7 - 6 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 89
7-10 If liquids other than water are See section 7.3.1.2 “Flushing the used as system liquid, flush with Liquid System”, 7-12 deionized water Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 7 - 7...
Call Illumina customer support especially Y-belt, for wear and to perform the task. replace faulty parts. Check parts for abrasion; wipe wear debris off, if necessary. 7 - 8 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 91
Please contact your local service organization, for further information on maintenance tasks and schedule of your system. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 7 - 9...
Leakages in the liquid system can also be caused by an empty liquid system or by aggressive liquids. When using aggressive liquids, take into account the chemical resistance of the tubing material. Refer to cross references above. 7 - 10 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 93
ATTENTION A leaking liquid system causes pipetting inaccuracy and cross-contamination. Never operate the Infinium LiHa and Infinium RoMa if the liquid system is leaking. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 7 - 11...
Page 94
ATTENTION Air bubbles in the liquid system cause pipetting inaccuracy. Never operate the Infinium LiHa and Infinium RoMa with air bubbles in the liquid system. 7 - 12 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 95
Place the tubing in a bottle with distilled or de-ionized water and flush the liquid system twice. Refer to cross references above. Flush the liquid system eight times with system liquid. Refer to cross references above. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 7 - 13...
Fig. 7-2 Syringe and valve Syringe screw Plunger lock screw If leakages still occur, replace the syringe or the syringe cap. Refer to cross references above. 7 - 14 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
If a tip is to be reinstalled, do not remove the lock nut from the tip. Always hold the tip at its upper end, avoiding contact with the coated surface whenever possible. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 7 - 15...
Page 98
Refer to cross references above. Reinstall the wash station on the worktable. Make sure that the wash station is pushed all the way back to the stop during installation. 7 - 16 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Contamination through waste liquid, if the containers are installed wrongly. Make sure not to mix up the system liquid container and the waste container. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 7 - 17...
Carriers and Racks Perform the following procedure to clean the carriers and racks. Remove all carriers and racks from the Infinium LiHa and Infinium RoMa worktable. The wash station can be cleaned on the worktable. Before cleaning, remove the barcode labels from the carriers, if possible.
Do not attempt to rotate the barcode scanner manually. Use the initialization routine to run the barcode scanner into maintenance position. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 7 - 19...
Page 102
Tube” sensor (B). Moisten a lint-free tissue with alcohol and clean the front surface of the “No Tube” sensor. Fig. 7-5 PosID “No Tube” sensor 7 - 20 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Note: Do not use alcohol or solvents to clean the arm guide. Do not use grease on the arm rails. Fig. 7-6 Arm guide and roller Arm guide Arm rail Arm guide roller Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 7 - 21...
Nucleic Acid solution. Residues Appropriate commercially available agents (e.g. DNAzap) are used to keep the pipetting area (worktable, carriers etc.) free of interfering nucleic acids. 7 - 22 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Which Errors The troubleshooting table below lists possible malfunctions and errors of the Infinium LiHa and Infinium RoMa. The operator is enabled to correct some of can the those problems or errors by him/herself. For this, appropriate corrective measures Operator are listed in the column “Corrective measures”.
Page 106
Put microplate on carrier Cannot pick up microplate Set gripper position Clean grippers Unusual noise during arm Worn or damaged parts Contact your local service organiza- movement tion 8 - 2 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Page 107
Algae block the wash sta- Clean the wash station tion Waste tubing kinked Check tubing for kinks Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 8 - 3...
Page 108
8 - Troubleshooting Troubleshooting Table 8 - 4 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
9 - Shutdown, Transport and Storage Shutdown 9 Shutdown, Transport and Storage Purpose of This This chapter instructs how to shut down the Infinium LiHa and Infinium RoMa, how to pack it for storage or transport, and specifies the storage and shipping Chapter conditions.
Page 110
Unplug the power cord from the power supply at the rear of the instrument. Fig. 9-3 RS232 interface on Optibo PCB RS232 interface connector USB connector Disconnect the instrument from the PC. 9 - 2 Infinium LiHa and Infinium RoMa Operating Manua -l Document # 1000000110155 v00...
Clean and, if necessary, decontaminate the entire instrument. 9.1.2 Reporting Fill out a copy of the decontamination form and place it with the instrument. Record the shut down in the appropriate CRM. Infinium LiHa and Infinium RoMa Operating Manua -l Document # 1000000110155 v00 9 - 3...
Illumina person only Transport The transport of the instrument may be done under direction of a qualified Illumina service person only. Due to the heavy weight, trained relocation staff is needed to lift the instrument.
Protect the instrument against dust and debris with a cover. For long term storage, pack the instrument in its original packing. Store all manuals and the “Maintenance and Service Logbook” with the instrument. Infinium LiHa and Infinium RoMa Operating Manua -l Document # 1000000110155 v00 9 - 5...
Page 114
9 - Shutdown, Transport and Storage Storage 9 - 6 Infinium LiHa and Infinium RoMa Operating Manua -l Document # 1000000110155 v00...
Product In accordance with the requirements specified in SJ/T11364-2014, all electronic and electrical Illumina products sold in the People’s Republic of China are labeled Marking with a marking for the restriction of the use of hazardous substances.
Page 116
10 - 2 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
Infinium LiHa and RoMa Accessory Kit Not for sale: KIT ACCESSORY INFINIUM ILLUMINA SCREWDRIVER 4.5/1.5*90 MM BRASS PLT.NI SCREWDRIVER SIZE 2 WORKTABLE MOUNT- SET WRENCH HEXAGON WRENCH HEXAGON 0.71 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 11 - 1...
11-2 landscape orientation ASSY ROMA ILLUMINA 150 mm (5.9 in.) a) Number or grid positions the carrier occupies Fig. 11-1 Carrier for 3 microplates landscape 11 - 2 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
TUBE 16MM 6 PCE. 25 mm 11-3 (0.98 in.) a) Number or grid positions the carrier occupies Fig. 11-2 Carrier for tubes (example for 16 tubes) Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 11 - 3...
1 waste position at center CAR. (0.98 in.) 8 wash positions deep at front a) Number or grid positions the carrier occupies Fig. 11-3 Wash/waste station, standard 11 - 4 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...
11 - Spare Parts and Accessories Tips and Accessories 11.5 Tips and Accessories Contact your local service representative for information on other accessories. Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00 11 - 5...
Page 122
11 - Spare Parts and Accessories Tips and Accessories 11 - 6 Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00...