MiSeq System Guide
Figure 20 Lower Sipper Handle
7
If the software does not identify the RFID of the PR2 bottle, see
8
Select Next.
Load the Reagent Cartridge
1
Open the reagent chiller door.
NOTE
Do not leave the reagent chiller door open for extended periods of time.
2
Hold the reagent cartridge on the end with the Illumina label, and slide the reagent cartridge into the
reagent chiller until the cartridge stops.
Always use the reagent cartridge associated with the type of flow cell that you loaded. If the reagent
cartridge is not compatible, a message appears on the screen. Select Back to load the appropriate
reagent cartridge or Home to return to the Home screen.
Figure 21 Load Reagent Cartridge
3
Close the reagent chiller door.
4
If the software does not identify the RFID of the reagent cartridge, see
page
44.
Document # 15027617 v04 Material # 20024228
For Research Use Only. Not for use in diagnostic procedures.
Resolve RFID Read Failure on page
Resolve RFID Read Failure on
44.