MiSeq System Guide
Rename Sample Sheets
1
From the Sample Sheets tab on the Manage Files screen, select from the following options:
Select the checkbox next to individual sample sheets.
u
Select the checkbox to the left of the Delete button to select all sample sheets in the list.
u
2
Select Rename.
3
Select the keyboard icon and use the on-screen keyboard to rename sample sheets.
4
Select Next.
5
Select Back.

Software Updates

If your system is connected to a network with internet access, you can automatically update the instrument
software from the Home screen. You can also configure the software to check automatically for BaseSpace
updates. For more information, see
If your instrument is not connected to a network with internet access, you can update the software manually.
Update Software Automatically
When software updates are available, the Update Available button appears on the Home screen. Otherwise,
this button is not visible. Make sure your MiSeq is connected to a network with internet access to enable this
option.
1
From the Home screen, select Update Available.
2
Confirm the command to update in the dialog box.
Reboot of the instrument is required. Installation of the update begins automatically upon reboot.
Update Software Manually
Use the Manual Update feature to update instrument control software and analysis software from the MiSeq
interface by browsing to the location of the installable software file.
1
From the Home screen, select Manage Instrument.
2
Select Software Update.
3
Select Browse to navigate to the location of the installable file for the new software version.
4
When the path to the installable software file appears on the screen, select Save and Update.
5
Confirm the command to update in the dialog box.
Reboot of the instrument is required. Installation of the update begins automatically upon reboot.

Shut Down the Instrument

It is best to leave the instrument on at all times. However, if the instrument must be turned off, use the
following procedure to shut down Windows and prepare the fluidics lines.
1
Perform a maintenance wash. For more information, see
2
Remove the waste bottle and discard the contents appropriately. Return the waste bottle to the reagent
compartment.
Document # 15027617 v04 Material # 20024228
For Research Use Only. Not for use in diagnostic procedures.
Configure Notifications of BaseSpace Updates on page
Perform a Maintenance Wash on page
12.
34.