Technical Specifications - Kendro HERAcell 240 Operating Instructions Manual

Co2 incubator with decontamination routine
Table of Contents

Advertisement

®
HERAcell
240 CO
Incubator Operating Instructions
2

13. Technical Specifications

Designation
Mechanical
Exterior dimensions (W x H x D)
Internal dimensions (W x H x D)
Sample chamber volume
Insert tray (W x D)
Number, standard delivery
Number, maximum
Surface load capacity, maximum
Total unit load, maximum
Weight, w./o. accessories
Thermal
Ambient temperature range
Temperature regulation range
Temperature deviation over time (DIN 12880, Part 2)
Decontamination temperature (ContraCon routine)
Temperature deviation, spatial (DIN 12880, part 2)
At 37 °C
auto-start routine duration, to 37 °C,
Ambient temperature: 20 °C
Temperature recovery time at 37 °C with door open for
30 seconds (to 98 % of the initial value)
Thermal loss to the environment:
At 37 °C
During ContraCon decontamination
Humidity
Water quality
Fill volumes:
Incubation mode
ContraCon decontamination mode
Constant humidity at 37 °C (high humidity mode)
Constant humidity at 37 °C (low humidity mode)
Humidity recovery time at 95 % rH, with door open for
30 seconds (to 98 % of the initial value)
1
The technical specifications cited above apply to units with a glass door.
Version: 03.2001 / 50065640
1
Unit of measure
mm
mm
mm
units
units
kg
kg
kg
kg
°C
°C
°C
°C
°C
hr.
min
kWh/h
kWh/h
ml
% rH
% rH
min
min
- 63 -
Value
780 x 934 x 834
607 x 670 x 583
l
approx. 238
560 x 500
3
12
10 / insert tray
30
approx. 85 (stainless steel)
approx. 90 (copper)
+18 ... 33
RT + 3 ... 55
± 0.1
90
± 0.5
5 ... 10
< 10
0.08
0.25
Distilled and sterile
l
4.5, max. / 1.8, min
350
approx. 95
approx. 90
approx. 18 (stainless steel)
approx. 18 (copper)

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Kendro HERAcell 240

Table of Contents