5.
System Description
Water Reservoir, Fig. 11:
The water reservoir [1] is integrated into the base of
the inner chamber and slopes towards the rear unit
wall. Bulges in the base of the pan serve as indicators
for the maximum fill level [3] and minimum fill level [2].
Rack System:
The support profiles for the rack systems have perfo-
rations every 42 mm. This allows the support brackets
to be inserted in locations to accommodate any cultu-
re vessel size. The insert trays have integrated tip-
protection and a stop to limit pull-out distance. Refer
to Chapter 6, "Commissioning", for a detailed descrip-
tion of the rack system.
Rear Unit Wall Openings, Fig. 12:
A pressure compensation opening [2] on the unit's re-
ar wall equalizes the pressure between the sample
chamber and the ambient environment.
An access port [1] which can be sealed when not in
use allows lines, tubes, or additional sensors to be
added to the unit's sample chamber.
NOTE –Operating conditions:
When auxiliary units are operated in the CO
incubator's sample chamber, the requirements of the
ambient conditions must be observed (refer to the ta-
ble). Energy brought into the sample chamber will in-
fluence the start of the temperature regulation range.
The inclusion of additional heat sources in the sample
chamber can result in the formation of condensation
(e.g., on the glass door).
Added energy Start of the temperature regulati-
on range
General
0 W
RT + 3 °C
5 W
RT + 6 °C
10 W
RT + 8.5 °C
15 W
RT + 11.5 °C
20 W
RT + 14 °C
25 W
RT + 17 °C
*RT = Room Temperature
Version: 03.2001 / 50065640
2
Example:
RT* = 21 °C
24 °C
27 °C
29.5 °C
32.5 °C
35 °C
38 °C
- 25 -