Setting The High/Low Humidity - Kendro HERAcell 240 Operating Instructions Manual

Co2 incubator with decontamination routine
Table of Contents

Advertisement

®
HERAcell
240 CO
Incubator Operating Instructions
2
7.
Control, Displays, Indicators
7.6

Setting the High/Low Humidity

No
Instruction
1. Activate the configuration
mode.
2. Display the mode.
3. Change the mode.
4. Save the target mode.
NOTE – Humidity presets:
The program controller has been factory-set to "high humidity".
Setting 0 (High) creates a relative humidity of approx. 95 % in the sample chamber.
Setting I (Low) a relative humidity of approx. 90 % is created.
Version: 03.2001 / 50065640
Operator action
Press the
key for 5 sec.
Press the
key.
Press the
key,
or
Press the
key.
Press the
key.
- 39 -
Display / explanation
All indicators on the control panel flash.
The current mode (High humidity) is shown
on the temperature display.
Example:
The new mode (Low humidity) is shown on
the temperature display.
Example:
The actual values are shown on the tempe-
rature and CO
displays.
2
Example:
The new mode is saved. The "low humidity"
mode is indicated by the yellow "Low Moisture
active" LED.

Hide quick links: