Error Code Overview - Kendro HERAcell 240 Operating Instructions Manual

Co2 incubator with decontamination routine
Table of Contents

Advertisement

®
HERAcell
240 CO
Incubator Operating Instructions
2
7.
Control, Displays, Indicators
7.9

Error Code Overview

Code
Designation
44
CO
metering cell faulty. NV RAM faulty
2
2
55
I
C bus error
66
Differences between the
individual temperature
probes.
77
CO
calibration range
2
exceeded.
88
Error during auto-start
99
Unit door(s) open.
100
Temperature below set-
point
101
Temperature above set-
point.
104
Temperature probe
faulty
200
CO
below setpoint
2
201
CO
above setpoint
2
204
CO
metering cell faulty. Sensor break/short circuit
2
500
ContraCon routine tem-
perature below setpoint
501
ContraCon routine tem-
perature above setpoint.
502
ContraCon routine error
Version: 03.2001 / 50065640
Possible cause
Measured value quality below
50 %.
The temperature signal is no
longer plausible.
Maximum compensation value
exceeded.
Total time has passed or the
max. compensation value has
been exceeded.
Door(s) open for longer than
10 minutes.net
Actual value < setpoint –1° C
Actual value > setpoint +1° C
Probe break / short circuit
Actual value < setpoint –1 %
No CO
;
2
Insufficient pressure;
Supply line blockage.
Actual value > setpoint +1 %
Pressure too high.
Actual value < 85° C
Actual value > 95° C
Power failure during the pre-
heat or hold phase.
- 43 -
Corrective action
Contact the Kendro Service.
Remove the source of the interference, e.g.,
cell phone.
Contact the Kendro Service.
Contact the Kendro Service.
Repeat auto-start.
Close unit door(s).
Contact the Kendro Service.
Make sure ambient temperature is below
permissible maximum. Contact the Kendro
Service.
Contact the Kendro Service.
Check the gas supply:
Change gas bottles.
Increase pressure to 1 bar, max.
Check the supply lines to the unit.
Check the gas supply:
Reduce pressure to 1 bar.
Contact the Kendro Service.
Repeat the decontamination routine. Contact
the Kendro Service as required.
Contact the Kendro Service.
Acknowledge the error by pressing the "Con-
traCon Routine" key (twice for 5 seconds),
then restart the routine.