Contracon Decontamination Routine Activation; Interrupting The Contracon Decontamination Routine - Kendro HERAcell 240 Operating Instructions Manual

Co2 incubator with decontamination routine
Table of Contents

Advertisement

®
HERAcell
240 CO
Incubator Operating Instructions
2
10. Cleaning and Disinfection

10.5 ContraCon Decontamination Routine Activation

No
Instruction
1. Switch the unit on.
2. Open both doors and wait
for the alarm signal to start
after 30 sec.
3. Start the ContraCon routine.
4. Close both unit doors.
5. End the ContraCon routine.

10.6 Interrupting the ContraCon Decontamination Routine

No
Instruction
1. Interrupt the ContraCon rou-
tine.
Version: 03.2001 / 50065640
Operator action
Press the main power switch. All displays on the control panel go on, the
Press the
key for 5 sec.
Press the
key for 5 sec.
Operator action
Press the
key for 5 sec.
- 53 -
Display / explanation
software version is shown on the temperature
display and on the am CO
The current actual values flash on the dis-
plays, the "Door" LED goes on, the time alarm
sounds after 30 sec.
The "ContraCon Routine" LED flashes.
The actual value appears on the temperature
display. The CO
display indicates the remai-
2
ning time. The "door" LED goes out.
Example:
The display switches back to the normal mo-
de (incubation mode).
Display / explanation
The routine jumps to the next phase in the
sequence. To completely abort the routine,
you must skip each phase individually by re-
peatedly pressing the key until the remaining
time shown is 0 h. You can then end the rou-
tine.
display.
2