10. Cleaning and Disinfection
Protective Screen and Fan Wheel Removal:
The fan wheel and its protective screen can be
disassembled if required.
1. Using the provided Allen wrench (3 mm), remove
the two fastening screws [3] on the protective
screen, then remove the screen.
2. The fan wheel [1] is secured to the shaft by the
stud screw [2]. Loosen the stud screw with the
Allan wrench (2 mm) and pull the fan wheel off the
shaft.
NOTE – Functional check:
After reinstallation, check the fan wheel to
make sure it is properly secured on the shaft and
turns freely. Reinstall the protective screen after this
check.
HERAcell
Figure 22
Fig. 22, Protective screen/fan wheel disassembly
- 48 -
®
240 CO
Incubator Operating Instructions
2
Version: 03.2001 / 50065640