11. Maintenance
11.8 Unit Fuse Replacement
The two identical unit fuses [4] are located in the fuse
box [1] next to the unit socket. The fuse characteri-
stics are:
Slo-blo fuse, 6.3 A ( 5 x 20 mm)
1. The fuse holder is secured in the fuse box [1] by
two clamp tabs [2].
2. To open, push the two tabs together and pull the
fuse holder [3] out of the fuse box.
3. Remove the blown fuse(s) from the fuse socket
and replace it/them with a new one.
4. Slide the fuse holder back into the fuse box and
gently press down on it until the clamp tabs lock in
place.

11.9 Door Seal Replacement

The door seal (magnetic seal) on the outer door rests
in a retainer groove. No tools are required to replace
the seal.
1. Pull the magnetic seal [3] out of the guide groove
[1].
2. Starting at one corner [2] insert the new seal into
the groove, then press the seal retainer strip [4]
into the retainer groove.
3. Make sure the tapered side of the retainer strip
sits properly in the retainer groove [1] and that the
seal is flush against the door frame.
HERAcell
Figure 27
Fig. 27, Unit fuse replacement
Fig. 28, Door seal replacement
- 60 -
®