Equalizing The Co Concentration - Kendro HERAcell 240 Operating Instructions Manual

Co2 incubator with decontamination routine
Table of Contents

Advertisement

11. Maintenance
11.6 Equalizing the CO
No
Instruction
1. Activate the equalization
function.
2. Display the setpoint.
3. Enter the measured value
(target value).
4. Save the target value.
5. Quit the equalization
process.
NOTE – Excess CO
An excessively high CO
approx. 30 seconds.
Concentration
2
Operator action
Press the
5 seconds
Press the
Press the
or
Press the
Press the
Press any key.
content after a CO
equalization:
2
2
content after equalization can be reduced by opening the doors for
2
HERAcell
key for
All indicators on the control panel flash.
key.
Preset value of 5 %
°C
Example:
key.
Target value, e.g.: 5.6 %
Example:
key.
"CAL" will appear briefly on the CO
key.
After this, the corrected actual value (measu-
red target value: 5.6 %) is displayed.
The actual values are shown on the tempe-
rature and the CO
- 58 -
®
240 CO
Incubator Operating Instructions
2
Display / explanation
displays.
2
Version: 03.2001 / 50065640
display.
2

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Kendro HERAcell 240

Table of Contents