7.10 Resetting the Overtemperature Protector

No
Instruction
1.
2. Switch the unit off.
3. Switch the unit on again
NOTE – Overtemperature protection:
Once the cause of the problem has been corrected (e.g., ambient temperature too high), the unit returns to
the normal incubation mode after being switched on again. If the cause cannot be corrected by simple measures
such as increasing the room ventilation or reducing the ambient temperature, the overtemperature protector will
immediately activate again. In this case, contact the Kendro Service.
Operator action
Press the main power switch.
Press the main power switch.
- 44 -
®
HERAcell
240 CO
Incubator Operating Instructions
2
Display / explanation
"Overtemperature protection" LED on.
All displays go out.
All displays on the control panel go on.
Version: 03.2001 / 50065640