11. Maintenance
11.7 Sterile Filter Replacement for the
CO

Gas Supply

2
The sterile filter has plastic threads and is screwed
into the socket on the circuit cabinet.
1. Make sure the gas supply is shut off.
2. Loosen the hose clamp [5].
3. Pull the gas hose [4] off the connecting piece on
the sterile filter [2].
4. Unscrew the retaining panel [1].
5. Unscrew the sterile filter [2] from its threaded sok-
ket [3].
6. Taking care not to cross-thread the new sterile
filter screw it into the socket ans hand-tighten.
7. Reinstall the retaining panel [1].
8. Insert the gas hose onto the sterile filter connec-
ting piece and secure it with the hose clamp.
Check to make sure the gas hose/filter connection
does not leak.
Version: 03.2001 / 50065640
Figure 26
Fig. 26, Sterile filter replacement
- 59 -