Support Bracket Installation:
1. Insert the support brackets [3] into the perfora-
tions [1] on the support profiles with the support
bar facing down.
2. Make sure the two vertical sections [2] of the sup-
port bracket lie against the support profile.
Inserting Trays:
1. Slide the insert trays [4] with their tip-protection [5]
towards the rear unit wall onto the support brak-
kets. The tip-protection [5] simultaneously serves
as the guide for the insert tray.
2. Slightly raise the insert tray so that the stop [6]
can be pushed over the support bracket.
3. Make sure the support bracket moves freely in the
two tip-protections.
Leveling the CO
Incubator:
2
1. Place a level on the middle insert tray.
2. Using the open-end wrench provided (24 mm),
turn the adjustable unit feet until the insert tray is
horizontal in all directions. Adjusting the height
with the feet should be performed from left to right
and from the rear to the front.
Version: 03.2001 / 50065640