Initial Setup; Sample Chamber Preparation; Rack System Installation; Fig. 13 Support Profile Installation/Removal - Kendro HERAcell 240 Operating Instructions Manual

Co2 incubator with decontamination routine
Table of Contents

Advertisement

6.

Initial Setup

6.1 Sample Chamber Preparation

The CO
incubator is not sterile at delivery. The unit
2
must be cleaned before beginning operation.
This applies to the following sample chamber compo-
nents:
Support profiles;
Support brackets;
Insert trays;
Inner incubator chamber;
Sample chamber seal;
Glass door.
NOTE – Cleaning:
Refer to Chapter 10 for detailed information
concerning cleaning and disinfecting the unit.
6.2

Rack System Installation

No tools are required to install the rack system. The
support profiles are held in place by spring pressure.
The support brackets are hung into the support pro-
files and the insert trays slide onto the support brak-
kets.
Support Profile Installation/Removal:
The sides of the support profiles are guided along
protrusions [2] and [5] and held in place by protrusions
[1] and [6]. The support profiles identified by the dia-
mond-shape (u) are installed against the rear unit
wall. The retainer spring clips [3] must face up.
1. Place the support profile [4] on the lower protrusi-
on [6] and swing it against the side sample cham-
ber wall so that it sits on protrusions [5] and [2].
2. Clip the spring clip [3] behind the upper protrusion
[1].
3. To remove the support profiles, pull the spring clip
down and out of the protrusion by its tab, then re-
move the support profile.
HERAcell
Figure 13
Fig. 13, Support profile installation/removal
- 26 -
®
240 CO
Incubator Operating Instructions
2
Version: 03.2001 / 50065640

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Kendro HERAcell 240

Table of Contents