3.
Location Requirements
3.6

Door Modifications

Both the exterior door and the glass door can be in-
stalled so that they open to the left or right. The door
stop can be retrofitted.
Standard versions of the incubator that have the full
glass door can be modified to have a 6-port divided
gas diaphragm. In this case, the gas diaphragm re-
places the glass door.
NOTE – Modifications:
Door modifications may only be performed by
the Kendro Laboratory Products Service.
3.7

Drawer

Once the unit has been set up, the drawer is inserted
between the two guide brackets under the unit body.
1. Completely open the exterior unit door to its stop.
2. Swing guide bracket [1] completely down.
3. Insert the drawer [4] with its raised edge [2] be-
hind the cable conduit [3]. The short instructions
[5] should face the front of the unit.
4. Slightly raise the front of the drawer and slide it
downwards and into place.
5. Push the drawer in to its pressure point, then
press down on the guide bracket and push it
completely back to its stop.
HERAcell
Figure 4
Fig. 4, Drawer installation
- 18 -
®
240 CO
Incubator Operating Instructions
2
Version: 03.2001 / 50065640