11. Maintenance
11.4 Equalizing the Temperature
No
Instruction
1. Activate the equalization
function.
2. Display the setpoint.
3. Enter the measured value
(target value).
4. Save the target value.
5. Quit the equalization
process.
Operator action
Press the
key for 5 se-
conds.
Press the
key.
°C
Press the
key.
or
Press the
key.
Press the
key.
Press any key.
- 56 -
®
HERAcell
240 CO
Incubator Operating Instructions
2
Display / explanation
All indicators on the control panel flash.
Preset value of 37 °C
Example:
Target value, e.g.: 36.4 °C
Example:
"CAL" will appear briefly on the temperature
display.
After this, the corrected actual value (measu-
red target value: 36.4 °C) is displayed.
The actual values are shown on the tempe-
rature and the CO
displays.
2
Version: 03.2001 / 50065640