Permobil M400 Corpus HD Service Manual page 92

Table of Contents

Advertisement

M400 Corpus HD
7.2.6.13 DIME Error
This occurs when the control system detects an ID conflict between
two modules in the system.
If a new module has been added:
• Disconnect the new module and switch power off and on.
• If no error occurs, connect the new module to the system and switch
power off and on.
• If the error recurs, the new module must be the cause of the
problem.
If no new modules have been added:
• Disconnect one module at a time and switch power off and on.
If the error persists after the checks listed above have been performed,
contact Permobil.
7.2.6.14 Memory Error
This is a non specific memory error that may be caused by any of the
system modules.
• Check all cables and connections.
• Switch power off and on.
If the error persists and the system includes third-party modules:
• Disconnect all modules that do not come from Penny & Giles Drives
Technology and switch power off and on.
If this has rectified the error:
• Connect one third-party module at a time and switch power off and
on each time.
• If the error recurs when the power is switched on, the last module to
be connected is defective.
If the error persists after the checks listed above, the power module may
be defective. Read more in 7.4 Repairing defective units , Page 98.
7.2.6.15 PM Memory Error
WARNING!
Incorrect programming may make the wheelchair
unsafe
Programming should only be performed by persons with knowledge of control
systems from Penny & Giles Drives Technology. Incorrect programming may
mean that the wheelchair is unsafe. Permobil cannot be held responsible for
losses of any kind if the control system factory settings are altered by
programming.
This is a specific error in the power module.
• Check all cables and connections.
• Reprogram the control system with the help of R-net PC
programmers.
This should be done with either the latest specific program file for the
wheelchair or the original Permobil program file.
92
Troubleshooting - Diagnostics R-net LCD control panel

Hide quick links: