Net-Total
Formulation/Second Tare Memory
Purpose
With this application program you can weigh in components for formulation of a mixture.
Preparation
Configure the Net-Total Formulation/Second Tare Memory application in the operating
menu: See "Configuring the Balance/Scale." Menu code: 2.1.3
Example
Step
1. Place an empty container on
the balance/scale
2. Tare the balance/scale
3. Add the first component
4. Store the first component weight.
If the print format is set to include
data ID codes, the following
is printed
5. Add the next component
6. Store the 2nd component weight
7. Add further components,
if desired
8. Display total weight and
fill to desired final weight
Key (or instruction) Display/Printout
+
+
t
+ 120,5
v
N1
+
v
As described
for steps 5 and 6
+ 191,0
c
65.0
g
0.0
g
g
0.0
g
NET
+
120.5 g
70,5
g
0.0
g
NET
g