Error Codes

Error codes are shown on the main display for 2 seconds. The program then returns automatically to the
previous mode (e.g., weighing).
Display
No segments appear
on the display
h
l
e 02
e 05
e 06
e 09
e 10
e 11
e 22
e 30
Max. weighing capacity The balance/scale was turned on
is less than indicated
under "Specifications"
The weight readout
is obviously wrong
If any other errors occur, contact your local Sartorius Service Center!
Cause
No AC power is available
The AC adapter is not plugged in
Battery or rechargeable
battery pack is discharged
The load exceeds the
balance/scale capacity
The weighing pan is not in place
Something is touching
the weighing pan
Calibration parameter
not met, e.g.:
– balance/scale not zeroed
– balance/scale is loaded
Selector is in "calibration/adjustment"
setting during normal weighing procedure
Built-in motorized calibration
weight is defective
Tare not possible when gross value < zero
The t key is blocked when there is data
in the second tare memory (net-total).
Only 1 tare function can be used at a time
Value input is not allowed for second
tare memory
Weight is too light or there is
no sample on the balance/scale
Interface port for
printer output is blocked
without the weighing pan in place
The balance/scale has not been
calibrated/adjusted
The balance/scale was not
tared before weighing
Solution
Check the AC power supply
Plug in the AC adapter
Replace the battery or recharge
the battery pack using an
external charger
Unload the balance/scale
Place the weighing pan on the
balance/scale
Move the object that is touching
the weighing pan
Calibrate only when zero
is displayed
Press t to tare the balance/scale
Unload the balance/scale
Return the selector to the basic
position
Contact your local
Sartorius Service Center
Tare the balance/scale
Press c to clear the tare
memory and release the tare key
Press t
Increase the weight on
the balance/scale
Contact your local
Sartorius Service Center
Place the weighing pan on
the balance/scale and press e
to turn the balance/scale back on
Calibrate/adjust the balance/scale
Tare the balance/scale
21