Weighing In Percent - Sartorius Talent Operating Instructions Manual

Electronic analytical and precision balances and precious metal scales with ec type approval
Hide thumbs Also See for Talent:
Table of Contents

Advertisement

Weighing in Percent

Purpose
This application program allows you to
obtain weight readouts in percent which are
in proportion to a reference weight.
Preparation
G Configure the Weighing in Percent applica-
tion in the operating menu:
See "Configuring the Balance/Scale."
1)
Menu code: 2. 1. 5
Example
Determine an unknown percentage: store the weight on the balance/scale as a reference percentage
Menu: Application program: Weighing in Percent (menu code 2. 1. 5)
Menu: Reference percentage: 100 % (menu code 3. 3. 5)
Step
1. Tare the balance/scale
2. Display the reference
percentage:
3. Place the reference weight for 100 %
on the balance/scale (here: 222.5 g)
4. Start the application; if the print
format is set to include data
ID codes the following is printed
5. Place an unknown weight
on the balance/scale (here: 322.5 g)
6. Display weight
7. Display percentage
8. Unload the balance/scale
9. Delete the reference percentage
10. Repeat the procedure starting from step 5, if desired.
$ Reference percentage:
Code 3. 3. 1
Code 3. 3. 2
Code 3. 3. 3
Code 3. 3. 4
Code 3. 3. 5
$ Storage parameter
(display accuracy for counting)
Code 3. 4. 1
Code 3. 4. 2
See also "Configuring the Balance/Scale"
Key (or instruction)
Display/Printout
t
v >2 sec.
+
+
v
Wxx%
+
+
v
+
v
c
1)
5 %
10 % (factory setting)
20 %
50 %
100 %
Standard resolution: With
stability (factory setting)
With 10 times higher
stability than standard
0.0
g
ref100
222.5
g
100.00
%
+
222.500 g
144.94
%
322.5
g
144.94
%
0.00
%
Not available in TE...PCE models
13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

GemGold

Table of Contents