ISO/GLP-compliant Printout

Features
You can have the parameters pertaining
to the ambient weighing conditions
printed before (GLP header) and after
(GLP footer) the values of a weighing
series.
These parameters include:
GLP header:
– Date
– Time at beginning of measurement
– Balance/scale manufacturer
– Balance/scale model
– Balance/scale serial number
– Software version number
– Identification number of the current
sampling operation
GLP footer:
– Date
– Time at end of measurement
– Field for operator signature
! The record can only be output to
a Sartorius data printer YDP03-0CE.
18
Settings
G Set the following menu codes
(see "Configuring the Balance/Scale"):
– GLP-compliant record: menu code 5 5 2
– ISO/GLP-compliant record after calibra-
tion/adjustment only:
menu code 7 2 2 or ISO/GLP-compliant
record always on: menu code 7 2 3
– Line format for printout: With data ID
codes – 22 characters: menu code 7 1 2
– Printout date/time:
– 24-h format: menu code 7 3 1
– 12-h format: menu code 7 3 2
! No ISO/GLP-compliant record is output
if any of the following settings are
configured:
menu codes 6 1 3, 6 1 4 (automatic
printout) and 7 1 1
Function Keys
Press p to output header and first
measured value.
> Header is output the first time p is
pressed
To output header and reference data
automatically with an application pro-
gram active: Press v
End application program:
End application program and output
GLP footer: Press c