About this Guide
This manual provides information on the Aurora flow cytometer, daily workflow, SpectroFlo™
software features, cytometer specifications, and instrument maintenance. It also includes
troubleshooting tips and service information.
Safety
Safety Symbols
The Aurora is intended for research use only. Not for diagnostic or therapeutic procedures.
The following table lists symbols used throughout this guide.
Symbol
Meaning
Caution: hazard or unsafe practice that could result in material damage,
data loss, minor or severe injury, or death
Risk of electric shock
Biological risk
General Safety
•
Do not place any object on top of the instrument.
•
Before turning on the cytometer, visually inspect all containers. Wear the recommended
protective laboratory attire such as protective gloves, eyewear, and lab coat.
•
Purge the sheath filter if air bubbles are visible in the sheath filter, or if the plenum or sheath
container have run dry.
•
Fill the sheath container as needed. Never use tap water as sheath solution.
•
Do not run bleach or detergent through the sheath filter. It is difficult to remove cleaning
solutions from the sheath filter.